Please use this identifier to cite or link to this item:
Title: Monte carlo simulation by code of MCNP and experimental check for measuring thickness of materials for the spec raltzrng system of MYO-101
Authors: Bui, Van Loat
Nguyen, Van Hung
Hoang, Sy Minh Phuong
Keywords: Monte-Carlo simulation;Monte Carlo N-Particle;Backscattering gamma;Scintillation detector;Nuclear technique
Issue Date: 2010
Publisher: H. : ĐHQGHN
Series/Report no.: Vol. 26;No. 1 (2010)
Abstract: At present, thickness measurement of materials based on effect of backscattering gamma-ray has been widely used in industry in our country. The report presented research in measuring thickness of some materials such as paper, plastic, aluminum and steel with using the dedicated system of MYO-101, having scintillation detector of YAP(Ce) and gamma-ray of 60 keV of Am-241 source, by Monte-Carlo simulation using the code of MCNP. The simulation was checked by experimental measurements. The results were shown that they were in accordance with the range of error. This research has been useful for training activities with a view of human resowces development in the field of application of nuclear techni.que in industry in Vietnam.
Description: p. 43-49
ISSN: 2588-1124
Appears in Collections:Mathematics and Physics

Files in This Item:

  • File : 117.pdf
  • Description : 
  • Size : 717.11 kB
  • Format : Adobe PDF

  • Items in DSpace are protected by copyright, with all rights reserved, unless otherwise indicated.